Lumiso Photometers – Advanced Optical Technology

How does Lumiso Work?

Lumiso photometers work, like all photometers, by shining light through a sample and measuring the amount of that light which passes through. The darker the colour of the sample, the less light will make it through to the detector.

Initially all wavelengths of visible light come out of the light source. These are reduced to a single wavelength using an optical filter. The photometer will measure the light passing through the sample in percentage transmission (%T), and this value is converted to a concentration using the internal calibration of the instrument. The optical platform refers to the light set-up which works to determine the amount of light transmitted through a sample.

What are the Main Features of the Lumiso Optical Platform?

Photometer performance and customer ease-of-use was at the forefront of our design for the Lumiso optical platform. Lumiso’s features and benefits include:

Improved accuracy between instruments

Lumiso photometers benefit from improved instrument calibration and chemistry calibrations that are certifiable to NIST traceable spectrophotometers. These advancements mean that measurement performance and accuracy between Lumiso instruments is improved.

New optical chamber design to allow for differences in glass diameter

The Lumiso optical chamber is made of compression rubber material to allow for differences in glass diameter. This also works to centre the test tube to the incident light beam shining through the sample, to ensure the most accurate results.

Streamlined optical chamber to make maintenance and cleaning easy

We have designed our new Lumiso instrument with the customer in mind. Dirt traps and tough-to-clean features have been removed so that the optics are easy to clean, ensuring performance accuracy is maintained.

How are Lumiso Photometers Calibrated?

Palintest calibration standards are produced by our quality control team and are traceable to national standards. Standards are characterised using a spectrophotometer which is checked using externally certified NIST standards before being used to assign values to calibration standards.

During the manufacturing and servicing of Lumiso, the instruments will be calibrated to read the correct %T at the wavelengths available on the instrument, using Palintest calibration standards.

Head Office

T: +44 (0)191 491 0808 

Palintest House
Kingsway
Team Valley
Gateshead
Tyne & Wear
NE11 0NS
United Kingdom

Australia Pacific

T: +61 1300 13 15 16

Palintest
53 Lorraine Street
Peakhurst Business Centre
Peakhurst
NSW 2210
Australia

Americas

T: +1 720 221 6878

Palintest
600 Corporate Circle
Suite F
Golden
Colorado
80401

USA